relation: http://acta.bibl.u-szeged.hu/85731/
title: Performance evaluation of a 3D-printed plastic V-groove nebulizer for trace analytical use in ICP-MS
creator:  Rweyemamu Almachiusi Rwegasira
creator:  Kajner Gyula
creator:  Bélteki Ádám
creator:  Ajtai Tibor
creator:  Cseh Martin
creator:  Geretovszky Zsolt
creator:  Galbács Gábor
subject: 01. Természettudományok
subject: 01.04. Kémiai tudományok
description: In this study, we demonstrated the feasibility and reliability of using 3D printing to create a low-cost, customized plastic V-groove nebulizer. Triplicates of plastic V-groove nebulizer were fabricated using multijet 3D-printing technology and tested for analytical performance using ICP-MS. The nebulizers were validated by optimizing key parameters and aerosol quality monitored by an optical particle counter (OPC). On average, all nebulizers generated aerosol droplets in the feasible range of diameters suitable for analytical use. Calibration curves for all nebulizers showed excellent linearity (r² > 0.998), with detection limits in the sub-ppb range.
publisher: University of Szeged
date: 2024
type: Konferencia vagy workshop anyag
type: NonPeerReviewed
format: part
language: hu
identifier: http://acta.bibl.u-szeged.hu/85731/1/proceedings_of_isaep_2024_262-265.pdf
identifier:    Rweyemamu Almachiusi Rwegasira;  Kajner Gyula;  Bélteki Ádám;  Ajtai Tibor;  Cseh Martin;  Geretovszky Zsolt;  Galbács Gábor:   Performance evaluation of a 3D-printed plastic V-groove nebulizer for trace analytical use in ICP-MS.   
language: eng