Kálomista Ildikó; Kéri Albert; Galbács Gábor: The effect of the plasma sampling depth and the flow rate of the aerosol dilution gas on the performance of single particle inductively coupled plasma mass spectrometry (SP-ICP-MS) measurements.
Előnézet |
Cikk, tanulmány, mű
proceedings_of_isaep_2015_349-353.pdf Letöltés (348kB) | Előnézet |
Absztrakt (kivonat)
Single Particle Inductively Coupled Plasma Mass Spectroscopy is a modern technique available for the characterization of nanoparticles. Optimization of the measurement conditions can increase the potential of the technique, especially with regards to the minimum detectable particle size (typically ca. 15 nm). To this end, in this work we explored the effects of changing the (interface) sampling depth and the on-line aerosol dilution by the High Matrix Introduction (HMI) function on an Agilent 7700X ICP-MS.
Mű típusa: | Konferencia vagy workshop anyag |
---|---|
Befoglaló folyóirat/kiadvány címe: | Proceedings of the International Symposium on Analytical and Environmental Problems |
Dátum: | 2015 |
Kötet: | 21 |
ISBN: | 978-963-306-411-5 |
Oldalak: | pp. 349-353 |
Konferencia neve: | International Symposium on Analytical and Environmental Problems (21.) (2015) (Szeged) |
Befoglaló mű URL: | http://acta.bibl.u-szeged.hu/55892/ |
Kulcsszavak: | Analitikai kémia |
Megjegyzések: | Bibliogr.: p. 352-353. ; összefoglalás angol nyelven |
Feltöltés dátuma: | 2018. dec. 10. 12:54 |
Utolsó módosítás: | 2022. aug. 08. 15:47 |
URI: | http://acta.bibl.u-szeged.hu/id/eprint/56000 |
Tétel nézet |